FDA Classifies Previously Announced Medtronic Physician Advisory On Small Subset Of Kappa And Sigma Pacemakers As Class I Recall
Main Category: Cardiovascular / CardiologyAlso Included In: Regulatory Affairs / Drug Approvals; Primary Care / General Practice; Public Health
Article Date: 12 Jun 2009 - 1:00 PDT
| Patient / Public: | ![]() | |
| Healthcare Prof: | ![]() |
Medtronic, Inc. (NYSE: MDT) announced the U.S. Food and Drug Administration (FDA) has classified Medtronic's previously announced physician advisory about a small subset of Kappa® and Sigma® series pacemakers, identified at increased susceptibility for separation of bonded interconnect wires from the electronic circuit, as a Class I recall. Notification to U.S. FDA, physicians and patients began May 18, 2009. As of the date of this news release, more than 95 percent of physicians following affected devices in the United States have confirmed receipt of the notification. The FDA's classification of Medtronic's Important Patient Safety Information does not alter the patient management recommendations already provided in the May 2009 letter, and physicians need not take any new action.
It is important to note that not all Medtronic Kappa and Sigma pacemakers are included in the advisory. Medtronic's May 2009 communication to physicians involves a small subset of approximately 2 percent of the 1.7 million Kappa and Sigma pacemakers implanted worldwide since 1997. Medtronic has identified an increased susceptibility for separated interconnect wires in a specific subgroup of 10,000 of these devices in the U.S. that remain in patients. Most of the devices in the potentially affected subset have been implanted for five years or longer and may be nearing normal elective replacement time.
In this limited subset of Kappa and Sigma series pacemakers, the issue can lead to loss of rate response, premature battery depletion, loss of telemetry, or no output, potentially causing patient symptoms such as fainting and lightheadedness. Patients who experience these symptoms are reminded to seek medical attention immediately.
Patients can determine if their device is included in the identified subset by entering their device serial number at http://www.KappaSigmaSNList.medtronic.com. Patients whose serial numbers match any of those found on the Web site should contact their physician to determine an appropriate course of action. Patients without Internet access, or who have these models of devices but are unsure if their devices are included in this advisory may also contact Medtronic Patient Services at 800-551-5544.
After consultation with the company's Independent Physician Quality Panel, Medtronic's May 2009 letter recommended that:
- Physicians should consider device replacement for patients who are both pacemaker dependent and who have been implanted with a device in the affected subsets.
- Physicians should continue routine follow up in accordance with standard practice for those patients who are not pacemaker dependent.
Source
Medtronic
Visit our cardiovascular / cardiology section for the latest news on this subject.
MLA
16 Feb. 2012. <http://www.medicalnewstoday.com/releases/153529.php>
APA
http://www.medicalnewstoday.com/releases/153529.php.
Please note: If no author information is provided, the source is cited instead.
|
Rate this article: (Hover over the stars then click to rate) |
Patient / Public: |
or |
Health Professional: |
Add Your Opinion
Please note that we publish your name, but we do not publish your email address. It is only used to let you know when your message is published. We do not use it for any other purpose. Please see our privacy policy for more information.
If you write about specific medications or operations, please do not name health care professionals by name.
All opinions are moderated before being included (to stop spam)
Contact Our News Editors
For any corrections of factual information, or to contact the editors please use our feedback form.
![]()
Please send any medical news or health news press releases to:
Note: Any medical information published on this website is not intended as a substitute for informed medical advice and you should not take any action before consulting with a health care professional. For more information, please read our terms and conditions.