New Breast Shield Reduces Radiation To Female Breast During Chest CT By Up To 73%

Main Category: Radiology / Nuclear Medicine
Article Date: 16 May 2005 - 9:00 PDT

email icon email to a friend   printer icon printer friendly   write icon opinions  


Current Article Ratings:

Patient / Public:4 stars

4 (1 votes)

Healthcare Prof:5 stars

5 (2 votes)


A new custom-designed tungsten-antimony breast shield reduces radiation dose to the female breast during chest CT by 43% to 73% without compromising diagnostic image quality, say researchers from the VCU Health System, Medical College of Virginia Hospitals and Physicians, in Richmond.

For the study, the researchers constructed a tissue-equivalent female breast, referred to as a breast "phantom," and measured the radiation dose incurred during a chest CT scan at different quadrants of the breast phantom, both with and without the use of their breast shield. They found their custom-designed breast shield effectively reduced the radiation dose delivered to different portions of the breast between 43% and 73%.

"Diagnostic chest CT is not performed to obtain diagnostic information about the breast, but rather about the lung tissue and mediastinum. The radiation dose delivered to the radiosensitive glandular tissues of the breast is an unwanted byproduct of their location. Few physicians are aware that both diagnostic chest CT and CT angiography imparts a radiation dose of 20 to 50 mGy or more to the breasts of an average-sized woman. That CT dose is roughly equivalent to 10 to 25 two-view mammograms and up to as many as 100 to 400 chest X-rays. The increasing radiation exposure from CT and its potential adverse effects are real concerns," said Mark S. Parker, MD, lead researcher on the study.

"When CT cannot be avoided, attempts are made to reduce the level of exposure, especially in young patients and females. Radiologists typically do this by manipulating the scanning protocol by reducing the tube current, the tube voltage or the exposure time or by increasing the table increment. However, each manipulation is associated with some compromise in image quality and, potentially, diagnostic information," said Dr. Parker.

According to the researchers, their radiation-absorbing, lead-free breast shield, which simply drapes across the patient, is lightweight, comfortable, reusable and could be used for CT angiography, as well as routine chest CT examinations. "All of our research has so far been conducted on breast and thorax tissue-equivalent phantoms. Our study is currently before the institutional review board at our institution to begin a trial on female patients. Based on the phantom studies, we are not anticipating significant image degradation. We hope to have the shield available for commercial use in the next several months," said Dr. Parker.

Dr. Parker will present the full results of the study on May 16 during the American Roentgen Ray Society Annual Meeting in New Orleans, LA.

The American Roentgen Ray Society (ARRS) was founded in 1900 and is the oldest radiology society in the U.S. Radiologists from all over the world attend the ARRS Annual Meeting to take part in instructional courses, scientific paper presentations, symposiums, new issues forums and scientific and commercial exhibits related to the field of radiology. The ARRS is named after Wilhelm Röentgen, who discovered the x-ray in 1895.

Contact: Jason Ocker
jason@arrs.org
703-858-4304
American Roentgen Ray Society
http://www.arrs.org

Article adapted by Medical News Today from original press release.
Visit our radiology / nuclear medicine section for the latest news on this subject.
There are no references listed for this article.
Please use one of the following formats to cite this article in your essay, paper or report:

MLA
Jason Ocker. "New Breast Shield Reduces Radiation To Female Breast During Chest CT By Up To 73%." Medical News Today. MediLexicon, Intl., 16 May. 2005. Web.
15 Feb. 2012. <http://www.medicalnewstoday.com/releases/24468.php>

APA
Jason Ocker. (2005, May 16). "New Breast Shield Reduces Radiation To Female Breast During Chest CT By Up To 73%." Medical News Today. Retrieved from
http://www.medicalnewstoday.com/releases/24468.php.

Please note: If no author information is provided, the source is cited instead.




Radiology / Nuclear Medicine

Most Popular Articles



Follow Our Radiology News On Twitter

Follow Us On Twitter
Get the latest news for this category delivered straight to your Twitter account. Simply visit our Radiology / Nuclear Medicine Twitter account and select the 'follow' option.



View list of all 'What Is...' articles »