Faster X-ray Interferometers Due To Single-photon Interference
Main Category: Radiology / Nuclear MedicineAlso Included In: Medical Devices / Diagnostics
Article Date: 23 Dec 2007 - 6:00 PDT
|
|
| Patient / Public: | ![]() |
5 (1 votes) |
| Healthcare Prof: | ![]() |
3 (1 votes) |
By means of X-ray interferometers, lengths down to the mm range can be measured with a resolution of less than one nm. The low translation velocity of the interferometers, which made their use in practice more difficult, could now be increased by a factor of 100 by exploiting the temporal correlation of singly interfering X-ray photons.
X-ray interferometers can measure lengths in the mm range with sub-nm resolution, whereby the almost perfect crystal grid of high-purity silicon is used as a length scale. The dimensions of any sub-µm-structured samples are thereby compared with the lattice parameter of silicon (ë0~0.543... nm) which has been determined very precisely within the scope of the project for the new definition of the Avogadro constant. For metrological applications in connection with scanning probe microscopes, such measurements are of great importance.
Up to now, a further spreading of this method had, however, been impeded by the low translation velocities of only 1 nm/s to 10 nm/s. They are due to the limited intensity of typical laboratory X-ray sources: the necessary filtering of the periodic interference signal leads to a reduction in contrast which, in a classic measurement, requires a slow translation of the interferometer.
In a quantum-mechanical sense, however, interference occurs also in a strongly "diluted" stream of X-ray photons: Regarded as a wave packet, even single photons follow in their temporal impact on the detector the same probability which, in the case of sufficiently intense X-ray light, leads to the continuous signal whose period one wants to determine. This well-known quantum-mechanical fact is now exploited for a specific purpose: if one protocols the times at which the single photons hit the detector, one can, by means of a subsequent Fourier transform of this time series, determine very precisely the frequency at which the lattice periods were passed. At constant velocity, it is then possible to reconstruct the path information, and one obtains the same information as with the classic measurement, but in a much shorter amount of time.
Thus, translation velocities of up to 1000 nm/s could be realised. This method will in future not only be used in further improved measuring arrangements for the determination of the lattice parameter of silicon, but also for other length measurements in nanotechnology.
###
Source:
U. Kuetgens
Physikalisch-Technische Bundesanstalt (PTB)
Visit our radiology / nuclear medicine section for the latest news on this subject.
MLA
14 Feb. 2012. <http://www.medicalnewstoday.com/releases/92296.php>
APA
http://www.medicalnewstoday.com/releases/92296.php.
Please note: If no author information is provided, the source is cited instead.
|
Rate this article: (Hover over the stars then click to rate) |
Patient / Public: |
or |
Health Professional: |
Add Your Opinion
Please note that we publish your name, but we do not publish your email address. It is only used to let you know when your message is published. We do not use it for any other purpose. Please see our privacy policy for more information.
If you write about specific medications or operations, please do not name health care professionals by name.
All opinions are moderated before being included (to stop spam)
Contact Our News Editors
For any corrections of factual information, or to contact the editors please use our feedback form.
![]()
Please send any medical news or health news press releases to:
Note: Any medical information published on this website is not intended as a substitute for informed medical advice and you should not take any action before consulting with a health care professional. For more information, please read our terms and conditions.